How can I set micropipette volume with decimal separators, for example 109.5 in p200?
$begingroup$
And how can I set:
- 150.1 in p200
- 871.3 in p1000
- 840.5 in p1000?
molecular-biology
$endgroup$
add a comment |
$begingroup$
And how can I set:
- 150.1 in p200
- 871.3 in p1000
- 840.5 in p1000?
molecular-biology
$endgroup$
$begingroup$
What is the brand and model of your pipet?
$endgroup$
– alec_djinn
Jan 17 at 6:51
add a comment |
$begingroup$
And how can I set:
- 150.1 in p200
- 871.3 in p1000
- 840.5 in p1000?
molecular-biology
$endgroup$
And how can I set:
- 150.1 in p200
- 871.3 in p1000
- 840.5 in p1000?
molecular-biology
molecular-biology
edited Jan 15 at 9:24
Laurel
1854
1854
asked Jan 14 at 18:09
SaraSara
141
141
$begingroup$
What is the brand and model of your pipet?
$endgroup$
– alec_djinn
Jan 17 at 6:51
add a comment |
$begingroup$
What is the brand and model of your pipet?
$endgroup$
– alec_djinn
Jan 17 at 6:51
$begingroup$
What is the brand and model of your pipet?
$endgroup$
– alec_djinn
Jan 17 at 6:51
$begingroup$
What is the brand and model of your pipet?
$endgroup$
– alec_djinn
Jan 17 at 6:51
add a comment |
3 Answers
3
active
oldest
votes
$begingroup$
It doesn't make sense to set volumes of decimals of microliters with a P200, since usually the accuracy of these pipettes is in the order of a few microliters. So even if you set your pipette at 190 uL, you may still dispense a volume ranging from 187 to 193 uL. For a P1000, due to higher volumes the inaccuracy is of course higher, usually +/- 10 uL or more. Some brands are more accurate than others, like Eppendorf or Finnpipette. I suggest you check the specs of your pipettes and see how much accuracy you can achieve.
$endgroup$
2
$begingroup$
I think a fair take-home message here is you shouldn't, and with many brands you often can't. Rainin and Eppendorf P1000's measure to four places from 0100 to 1000, for example. The best bet is to use two different pipets, and use them in that fashion throughout the process to maintain consistency.
$endgroup$
– CKM
Jan 14 at 21:10
add a comment |
$begingroup$
I would always use multiple pipetters in cases like that, but really the answer is that no pipetter is going to give you 4 sig figs of accuracy. It's just not possible with the technology.
$endgroup$
add a comment |
$begingroup$
For p200 the smallest unit is the tenth of a ul (0.1) but the intervals go by 0.2 so to set 109.5 (assuming your pipette can be adjusted between intervals) you need to set the pipette so the pointer is equidistant from the 109.4 and 109.6 mark. For the p1000 the smallest unit is 1 ul so you cannot set it to 871.3.
$endgroup$
$begingroup$
Some brands of pipeters "click" at each allowed increment, so you won't be able to go between them.
$endgroup$
– swbarnes2
Jan 14 at 23:10
$begingroup$
Interesting. Thanks for sharing I didn't know that.
$endgroup$
– Cell
Jan 15 at 1:53
add a comment |
Your Answer
StackExchange.ifUsing("editor", function () {
return StackExchange.using("mathjaxEditing", function () {
StackExchange.MarkdownEditor.creationCallbacks.add(function (editor, postfix) {
StackExchange.mathjaxEditing.prepareWmdForMathJax(editor, postfix, [["$", "$"], ["\\(","\\)"]]);
});
});
}, "mathjax-editing");
StackExchange.ready(function() {
var channelOptions = {
tags: "".split(" "),
id: "375"
};
initTagRenderer("".split(" "), "".split(" "), channelOptions);
StackExchange.using("externalEditor", function() {
// Have to fire editor after snippets, if snippets enabled
if (StackExchange.settings.snippets.snippetsEnabled) {
StackExchange.using("snippets", function() {
createEditor();
});
}
else {
createEditor();
}
});
function createEditor() {
StackExchange.prepareEditor({
heartbeatType: 'answer',
autoActivateHeartbeat: false,
convertImagesToLinks: false,
noModals: true,
showLowRepImageUploadWarning: true,
reputationToPostImages: null,
bindNavPrevention: true,
postfix: "",
imageUploader: {
brandingHtml: "Powered by u003ca class="icon-imgur-white" href="https://imgur.com/"u003eu003c/au003e",
contentPolicyHtml: "User contributions licensed under u003ca href="https://creativecommons.org/licenses/by-sa/3.0/"u003ecc by-sa 3.0 with attribution requiredu003c/au003e u003ca href="https://stackoverflow.com/legal/content-policy"u003e(content policy)u003c/au003e",
allowUrls: true
},
noCode: true, onDemand: true,
discardSelector: ".discard-answer"
,immediatelyShowMarkdownHelp:true
});
}
});
Sign up or log in
StackExchange.ready(function () {
StackExchange.helpers.onClickDraftSave('#login-link');
});
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Post as a guest
Required, but never shown
StackExchange.ready(
function () {
StackExchange.openid.initPostLogin('.new-post-login', 'https%3a%2f%2fbiology.stackexchange.com%2fquestions%2f80521%2fhow-can-i-set-micropipette-volume-with-decimal-separators-for-example-109-5-in%23new-answer', 'question_page');
}
);
Post as a guest
Required, but never shown
3 Answers
3
active
oldest
votes
3 Answers
3
active
oldest
votes
active
oldest
votes
active
oldest
votes
$begingroup$
It doesn't make sense to set volumes of decimals of microliters with a P200, since usually the accuracy of these pipettes is in the order of a few microliters. So even if you set your pipette at 190 uL, you may still dispense a volume ranging from 187 to 193 uL. For a P1000, due to higher volumes the inaccuracy is of course higher, usually +/- 10 uL or more. Some brands are more accurate than others, like Eppendorf or Finnpipette. I suggest you check the specs of your pipettes and see how much accuracy you can achieve.
$endgroup$
2
$begingroup$
I think a fair take-home message here is you shouldn't, and with many brands you often can't. Rainin and Eppendorf P1000's measure to four places from 0100 to 1000, for example. The best bet is to use two different pipets, and use them in that fashion throughout the process to maintain consistency.
$endgroup$
– CKM
Jan 14 at 21:10
add a comment |
$begingroup$
It doesn't make sense to set volumes of decimals of microliters with a P200, since usually the accuracy of these pipettes is in the order of a few microliters. So even if you set your pipette at 190 uL, you may still dispense a volume ranging from 187 to 193 uL. For a P1000, due to higher volumes the inaccuracy is of course higher, usually +/- 10 uL or more. Some brands are more accurate than others, like Eppendorf or Finnpipette. I suggest you check the specs of your pipettes and see how much accuracy you can achieve.
$endgroup$
2
$begingroup$
I think a fair take-home message here is you shouldn't, and with many brands you often can't. Rainin and Eppendorf P1000's measure to four places from 0100 to 1000, for example. The best bet is to use two different pipets, and use them in that fashion throughout the process to maintain consistency.
$endgroup$
– CKM
Jan 14 at 21:10
add a comment |
$begingroup$
It doesn't make sense to set volumes of decimals of microliters with a P200, since usually the accuracy of these pipettes is in the order of a few microliters. So even if you set your pipette at 190 uL, you may still dispense a volume ranging from 187 to 193 uL. For a P1000, due to higher volumes the inaccuracy is of course higher, usually +/- 10 uL or more. Some brands are more accurate than others, like Eppendorf or Finnpipette. I suggest you check the specs of your pipettes and see how much accuracy you can achieve.
$endgroup$
It doesn't make sense to set volumes of decimals of microliters with a P200, since usually the accuracy of these pipettes is in the order of a few microliters. So even if you set your pipette at 190 uL, you may still dispense a volume ranging from 187 to 193 uL. For a P1000, due to higher volumes the inaccuracy is of course higher, usually +/- 10 uL or more. Some brands are more accurate than others, like Eppendorf or Finnpipette. I suggest you check the specs of your pipettes and see how much accuracy you can achieve.
answered Jan 14 at 19:06
betelgeusebetelgeuse
1062
1062
2
$begingroup$
I think a fair take-home message here is you shouldn't, and with many brands you often can't. Rainin and Eppendorf P1000's measure to four places from 0100 to 1000, for example. The best bet is to use two different pipets, and use them in that fashion throughout the process to maintain consistency.
$endgroup$
– CKM
Jan 14 at 21:10
add a comment |
2
$begingroup$
I think a fair take-home message here is you shouldn't, and with many brands you often can't. Rainin and Eppendorf P1000's measure to four places from 0100 to 1000, for example. The best bet is to use two different pipets, and use them in that fashion throughout the process to maintain consistency.
$endgroup$
– CKM
Jan 14 at 21:10
2
2
$begingroup$
I think a fair take-home message here is you shouldn't, and with many brands you often can't. Rainin and Eppendorf P1000's measure to four places from 0100 to 1000, for example. The best bet is to use two different pipets, and use them in that fashion throughout the process to maintain consistency.
$endgroup$
– CKM
Jan 14 at 21:10
$begingroup$
I think a fair take-home message here is you shouldn't, and with many brands you often can't. Rainin and Eppendorf P1000's measure to four places from 0100 to 1000, for example. The best bet is to use two different pipets, and use them in that fashion throughout the process to maintain consistency.
$endgroup$
– CKM
Jan 14 at 21:10
add a comment |
$begingroup$
I would always use multiple pipetters in cases like that, but really the answer is that no pipetter is going to give you 4 sig figs of accuracy. It's just not possible with the technology.
$endgroup$
add a comment |
$begingroup$
I would always use multiple pipetters in cases like that, but really the answer is that no pipetter is going to give you 4 sig figs of accuracy. It's just not possible with the technology.
$endgroup$
add a comment |
$begingroup$
I would always use multiple pipetters in cases like that, but really the answer is that no pipetter is going to give you 4 sig figs of accuracy. It's just not possible with the technology.
$endgroup$
I would always use multiple pipetters in cases like that, but really the answer is that no pipetter is going to give you 4 sig figs of accuracy. It's just not possible with the technology.
answered Jan 15 at 16:56
swbarnes2swbarnes2
3,555811
3,555811
add a comment |
add a comment |
$begingroup$
For p200 the smallest unit is the tenth of a ul (0.1) but the intervals go by 0.2 so to set 109.5 (assuming your pipette can be adjusted between intervals) you need to set the pipette so the pointer is equidistant from the 109.4 and 109.6 mark. For the p1000 the smallest unit is 1 ul so you cannot set it to 871.3.
$endgroup$
$begingroup$
Some brands of pipeters "click" at each allowed increment, so you won't be able to go between them.
$endgroup$
– swbarnes2
Jan 14 at 23:10
$begingroup$
Interesting. Thanks for sharing I didn't know that.
$endgroup$
– Cell
Jan 15 at 1:53
add a comment |
$begingroup$
For p200 the smallest unit is the tenth of a ul (0.1) but the intervals go by 0.2 so to set 109.5 (assuming your pipette can be adjusted between intervals) you need to set the pipette so the pointer is equidistant from the 109.4 and 109.6 mark. For the p1000 the smallest unit is 1 ul so you cannot set it to 871.3.
$endgroup$
$begingroup$
Some brands of pipeters "click" at each allowed increment, so you won't be able to go between them.
$endgroup$
– swbarnes2
Jan 14 at 23:10
$begingroup$
Interesting. Thanks for sharing I didn't know that.
$endgroup$
– Cell
Jan 15 at 1:53
add a comment |
$begingroup$
For p200 the smallest unit is the tenth of a ul (0.1) but the intervals go by 0.2 so to set 109.5 (assuming your pipette can be adjusted between intervals) you need to set the pipette so the pointer is equidistant from the 109.4 and 109.6 mark. For the p1000 the smallest unit is 1 ul so you cannot set it to 871.3.
$endgroup$
For p200 the smallest unit is the tenth of a ul (0.1) but the intervals go by 0.2 so to set 109.5 (assuming your pipette can be adjusted between intervals) you need to set the pipette so the pointer is equidistant from the 109.4 and 109.6 mark. For the p1000 the smallest unit is 1 ul so you cannot set it to 871.3.
edited Jan 15 at 1:55
answered Jan 14 at 21:17
CellCell
385116
385116
$begingroup$
Some brands of pipeters "click" at each allowed increment, so you won't be able to go between them.
$endgroup$
– swbarnes2
Jan 14 at 23:10
$begingroup$
Interesting. Thanks for sharing I didn't know that.
$endgroup$
– Cell
Jan 15 at 1:53
add a comment |
$begingroup$
Some brands of pipeters "click" at each allowed increment, so you won't be able to go between them.
$endgroup$
– swbarnes2
Jan 14 at 23:10
$begingroup$
Interesting. Thanks for sharing I didn't know that.
$endgroup$
– Cell
Jan 15 at 1:53
$begingroup$
Some brands of pipeters "click" at each allowed increment, so you won't be able to go between them.
$endgroup$
– swbarnes2
Jan 14 at 23:10
$begingroup$
Some brands of pipeters "click" at each allowed increment, so you won't be able to go between them.
$endgroup$
– swbarnes2
Jan 14 at 23:10
$begingroup$
Interesting. Thanks for sharing I didn't know that.
$endgroup$
– Cell
Jan 15 at 1:53
$begingroup$
Interesting. Thanks for sharing I didn't know that.
$endgroup$
– Cell
Jan 15 at 1:53
add a comment |
Thanks for contributing an answer to Biology Stack Exchange!
- Please be sure to answer the question. Provide details and share your research!
But avoid …
- Asking for help, clarification, or responding to other answers.
- Making statements based on opinion; back them up with references or personal experience.
Use MathJax to format equations. MathJax reference.
To learn more, see our tips on writing great answers.
Sign up or log in
StackExchange.ready(function () {
StackExchange.helpers.onClickDraftSave('#login-link');
});
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Post as a guest
Required, but never shown
StackExchange.ready(
function () {
StackExchange.openid.initPostLogin('.new-post-login', 'https%3a%2f%2fbiology.stackexchange.com%2fquestions%2f80521%2fhow-can-i-set-micropipette-volume-with-decimal-separators-for-example-109-5-in%23new-answer', 'question_page');
}
);
Post as a guest
Required, but never shown
Sign up or log in
StackExchange.ready(function () {
StackExchange.helpers.onClickDraftSave('#login-link');
});
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Post as a guest
Required, but never shown
Sign up or log in
StackExchange.ready(function () {
StackExchange.helpers.onClickDraftSave('#login-link');
});
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Post as a guest
Required, but never shown
Sign up or log in
StackExchange.ready(function () {
StackExchange.helpers.onClickDraftSave('#login-link');
});
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Post as a guest
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
$begingroup$
What is the brand and model of your pipet?
$endgroup$
– alec_djinn
Jan 17 at 6:51